Newton Scientific Fume Hood
Newton Scientific ducted fume hood protect user from being exposed to toxic fumes and hazardous particulate by pulling out contaminated air via the building’s HVAC system and out of the building. This allows contaminated air inside the laboratory to be pushed out from the building and ensure polluted air is not recirculated back into the laboratory.
Features
- Double layered design
- Aerodynamic front fascia
- Industrial grade galvanised steel construction
- Heavy epoxy powder coated finish to withstand harsh laboratory environments
- Phenolic resin inner liner and baffle with built in slots for efficient scavenging
- Dished epoxy resin worktop with high chemical and heat resistance, black
- Purpose built light box external to airflow, preventing any corrosion and minimizing turbulence
- Electronic light ballast for flicker free usage and user eye comfort
- Counterbalanced sash assembly for years of trouble free use
- Tempered safety glass with Easygrip® hand rail
- 2 side access panels built into inner liner for future addition of utilities
- Stainless steel airfoil designed to maximize air sweep of work surface
- All fume hoods comply to exhaust requirements and face velocity test as per ASHRAE 110-95; when ordered with control systems, units also comply to BS 7258 and EN 14175
Product Specification
Model | 4 Feet, Ducted | 5 Feet, Ducted | 6 Feet, Ducted | 8 Feet, Ducted |
Cat No. | 61000660 | 61000661 | 61000662 | 61000663 |
Dimensions (LxWxH), Including base cabinet | 1220 mm x 920mm x 2420mm | 1520mm x 920mm x 2420mm | 1820mm x 920mm x 2420mm | 2420mm x 920mm x 2420mm |
Body | Electrogalvanised steel with heavy epoxy powder coated finish | |||
Worktop | Dished epoxy resin, black with marine edge | |||
Inner Liner & Baffle | Phenolic resin, laboratory grade | |||
Airfoil | SS 304 | |||
Standards Complied | • ASHRAE 110-95 | |||
• BS 7258 (with Optional Controller) | ||||
• EN 14175 (with Optional Controller) | ||||
Services | • 1 Pre Plumbed Cold Water Remote | |||
• 1 Pre plumbed Gas Remote | ||||
• 2 x Pre wired 13 Amp Switch socket outlets (BS) | ||||
• PP Cup Sink | ||||
Controls | Fan ON/OFF switch | |||
Light ON/OFF switch | ||||
Base | Epoxy powder coated steel construction | |||
Duct Collar | 200 mm | 250 mm | 300 mm | 300 mm x 2 |
Design Velocity | 0.5 m/s at 500mm sash height | |||
Extract Volume at 500mm Height Sash Opening | 1188 CMH | 1512 CMH | 1836 CMH | 2481 CMH |
Top View | Side View | Front View |
|
|
|